CWA ID92453. Will Not Scan With RF Errors.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2254278 3T RF Amplifier Enable Line fault 93. The RF amplifier indicates that it was unblanked continuously for an excessive amount of time. Error: 2254278 Host: CENTMR02 Process: NSP File: SCP:../rf_nb_contro Line: 1491 3T RF Amplifier Enable Line (fault 93). The RF amplifier indicates that it was unblanked continuously for an excessive amount of time. Check PSD parameters. If the fault persists, replace the RF amplifier.

Applies to

Models
Excite
Product
3T Excite HD 12.0 software only
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution