Pular para o conteúdo principal
GE Healthcare · Nuclear MedicineInfinia

Hawkeye offsets do not stabilize, CT session fails to start (User Message

Symptom

Hawkeye offsets do not stabilize, CT session fails to start (User Message: Cannot prepare data needed for X-ray image correction)

Error message

CT_ERROR_AIA_OFFSET,6105,ERROR:.: Error AIA could not stabilize offsets.:: Cannot prepare data needed for Xray image correction

CT_ERROR_OFFSET_STABILITY,6106,ERROR.:: Error offsets stability check failed.:: Error in CT Device

Applies to

Models
Infinia
Product
Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software)
Component
*X-Ray (Generator, DAS, Controller) *X-Ray (Generator, DAS, Controller) 5165340 DAS Converter Board Hawkeye 4 5165340 DAS Converter Board Hawkeye 4 5165340 DAS Converter Board Hawkeye 4
Area
Nuclear Medicine