Symptom

GT table would reset when cradle is driven. Up and down work without error See Long Description for more details GT table would reset when driven. Two amber led's lit on the GTCB, cradle fault and ims fault. Up and down worked.

Error message

Error: 260134602 Cradle drive torque setting error See Long Description for error example Mon Dec 26 10:41:58 2005 Error: 260134602 Host: tgp Process: tgp_subsys File: gtcb_cmc_drv_req.cxx Line: 928 Mon Dec 26 10:41:58 2005 Host : Table Sub-System Ermes # : 260134602 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_drv_req.cxx Line# : 928 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle drive torque setting error

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5306599 (replaces 5132669-2 and 5132669-3) Step Motor Driver Positioning GT P - used on GT and VT tables See Long Description for details Part #: 5306599 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Comments: Per Kittyhawk RSPL, 212008898, 6/9/08 Part #: 5132669-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Part #: 5132669-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: STEP MOTOR DRIVER Positioning GT FRU: Yes Status: Orderable Comments: added per to PET Table RSPL. source pole/planner code revised per to Global Table transfer
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.