Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed2601181775005450148260118362956559655956559630

BrightSpeed HPower, Delight, Velocity or MERI HPower

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118177 Jedi (40-0114)Rotor not at speed. See Long Description for complete error Thu Feb 15 14:07:54 2007 260118177 4 orp orp_subsys gen_exec.c 2376 Thu Feb 15 14:07:54 2007 Host : Jedi Ermes # : 260118177 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Static Scan: 50054/1/1 (40-0114) Rotor not at speed . Z6,Z7= 2h,0h Mon Apr 17 14:01:27 2006 Error: 260118177 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Mon Apr 17 14:01:26 2006 Host : Jedi Ermes # : 260118177 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50148/1/1 (40-0114)Rotor not at speed. ^C Z6,Z7= 2h,0h

Error: 260118362 Rotor Event Timestamps See Long Description for details Mon Jan 22 15:11:07 2007 Error: 260118362 Host: orp Process: orp_subsys File: gen_exec.c Line: 484 Mon Jan 22 15:11:06 2007 Host : On-Board Rotating Processor Ermes # : 260118362 Exception Class : Secondary Severity : Sec/Soft File : gen_exec.c Line# : 484 Function : No System Function Reported Scan Type : Axial Scan: 4987/2/1 Rotor Event Timestamps(mS) Cur Time: 95

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5128609-7 (replaces 5128609) Anode HV CABLE Used on HP60 and BrightSpeed Systems. Introduced April 2006 to correct thermal failure of 2241313-2 P9230TG High Voltage Anode cable See Long Description for details 5128609-7 HV cable - Anode Part #: 5128609-7 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLE FRU: Yes Status: Orderable Old FRU Compatible No Comments Added per to ECR2115693 by 305024364 in April 7, 2011 Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5128609-7 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Direct ECO Number ECR2115693 Part #: 5128609 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLE Positioning HP60 FRU: Yes Status: Orderable Comments: do not place new order for old p/n(2241313-2), existing old part should be used on HiSpeed Zx/i series only.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues