Pular para o conteúdo principal
GE Healthcare · MRSigna2225669

ERBTEC RF Amplifier fault 90 Due to Defective IPA Tube.

Symptom

Unable to scan with Erbtec errors

Error message

2225669 ERBTEC RF Amplifier fault 90. IPA idle current fault. The IPA tube cannot produce any idle current. Suite: MRS1 Host: OC0 Proc: HPC PORT ID Error: 2225669 File: PMH.m 1.29 Method: spi rf mon Line: 3035 ERBTEC RF Amplifier fault 90. IPA idle current fault. The IPA tube cannot produce any idle current. Possible defective IPA tube, bad cable, power control board, or bad A/D on processor board.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
46-305000P3 ERBTEC 176–1001 3CPX800A7 TUBE (IPA Tube)
Area
MR

Full cause & step-by-step resolution