GE Healthcare · MRSigna2225669
ERBTEC RF Amplifier fault 90 Due to Defective IPA Tube.
Symptom
Unable to scan with Erbtec errors
Error message
2225669 ERBTEC RF Amplifier fault 90. IPA idle current fault. The IPA tube cannot produce any idle current. Suite: MRS1 Host: OC0 Proc: HPC PORT ID Error: 2225669 File: PMH.m 1.29 Method: spi rf mon Line: 3035 ERBTEC RF Amplifier fault 90. IPA idle current fault. The IPA tube cannot produce any idle current. Possible defective IPA tube, bad cable, power control board, or bad A/D on processor board.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-305000P3 ERBTEC 176–1001 3CPX800A7 TUBE (IPA Tube)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.