Pular para o conteúdo principal
GE Healthcare · CTLightSpeed186218

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 186218 detail error code : 40 - 0114H Rotation error : HV cable short circuit error See LD for complete error Mon Jun 12 11:08:29 2006 Error: 186218 Host: BAYCT Process: tgp_out File: tgp_in_input.c Line: 324 XG Error phase is 3 code is 40. detail error code : 40 - 0114H Rotation error : HV cable short circuit error

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5128609-7 (replaces 5128609) Anode HV CABLE Used on HP60 and BrightSpeed Systems. Introduced April 2006 to correct thermal failure of 2241313-2 P9230TG High Voltage Anode cable See Long Description for details 5128609-7 HV cable - Anode Part #: 5128609-7 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLE FRU: Yes Status: Orderable Old FRU Compatible No Comments Added per to ECR2115693 by 305024364 in April 7, 2011 Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5128609-7 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Direct ECO Number ECR2115693 Part #: 5128609 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLE Positioning HP60 FRU: Yes Status: Orderable Comments: do not place new order for old p/n(2241313-2), existing old part should be used on HiSpeed Zx/i series only.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.