Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT

92305 LightSpeed 1.X

Symptom

After software upgrade to 06MW29.7, the scanner clock CHANGED Forward by few hours after reboot

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
06MW29.7 - supports HS QX/I, LS 1X, 2X, 3X, 4X, 5X, BrightSpeed & 7X GRE host consoles, except LS 5X RT16/Xtra See Long Description for more information 06MW29.7 - supports HS QX/I, LS 1X, 2X, 3X, 4X, 5X, BrightSpeed & 7X GRE host consoles, except LS 5X RT16/Xtra CLASSM & CLASSC 5181129 = Class M for 06MW29.7 5192235 = Class C for 06MW29.7 Throw away 06MW29.7 Class C CD, p/n 5182628 ! ! ! replace it with Class C CD, p/n 5192235 REMOTE QUERY OF IG, VIG, DARC2, & VDARC TO DETERMINE IF “WESTVILLE” OR “JARRELL” VERSION · “Westville” computer in IG, VIG, DARC2, & VDARC end of life · IG, VIG, DARC2, VDARC subassemblies use the same computer, whether “Westville” or “Jarrell” · The current Intel “Westville” computer is being replaced by Intel’s “Jarrell” · “Jarrell” (IG, VIG, DARC2, VDARC) targeted for mfg cut-in in Milwaukee & Hino FW39-40’06; Beijing later · “dmidecode” command may be used remotely to determine which is installed in the host console 1. Login to IG, VIG, DARC2, or VDARC as “root” 2. Type dmidecode | less 3. View “Version” line towards top of response If SWV25 , then WESTVILLE installed If SE7520 JR 23 , then JARRELL installed NOTE: In multi-IG configuration where both Westville & Jarrell are installed in same host console (mixed config), the Jarrell IG must be in Slot #1 to optimize speed/performance. WESTVILLE RESPONSE: (all subassemblies use the same Intel Westville computer) $ su - $ dmidecode | less # dmidecode 2.6 # dmidecode 2.2 SMBIOS 2.3 present. 73 structures occupying 2782 bytes. Table at 0x000F1260. Handle 0x0000 DMI type 0, 20 bytes. BIOS Information Vendor: Intel Corporation Version: SWV25 .86B.0257.P42.0504011224 Release Date: 04/01/2005 …etc…. JARRELL RESPONSE: (all subassemblies use the same Intel Jarrell computer) $ su - $ dmidecode | less # dmidecode 2.6 SMBIOS 2.3 present. 59 structures occupying 2536 bytes. Table at 0x000FD4A0. Handle 0x0000 DMI type 0, 20 bytes. BIOS Information Vendor: Intel Corporation Version: SE7520 JR 23 .86B.P.10.00.0087.120820051348 Release Date: 12/08/2005 …etc…. “Jarrell” SOFTWARE RELEASE 06MW29.7 CLASS C AND M TAB #182-DISTRIBUTED 19SEP06 “Jarrell” software release 06MW29.7 Class C & M CDs TAB #182-distributed 19-Sep-06 5181129 = Class M for 06MW29.7 5182628 = Class C for 06MW29.7 - superceded by 06MW29.7 Class C, p/n 5192235 throw away 06MW29.7 Class C CD, p/n 5182628 replace it with Class C CD, p/n 5192235 -> will be TAB #182-distributed with cover letter Class C p/n 5182628, GRE diagnostics sw, incorrectly reads VIG/VRAC & VDARC/VDIP identification data It sees a mismatch between the “expected” and “actual” ID string read from the device The word “Class” is inserted in the “actual” ID string Given this mismatch, it reports a hardware FAILURE when hardware is “good” This issue is fixed on Class C, p/n 5192235 The following steps may be used to determine if the correct Class C software (p/n 5192235) is loaded: 1. Type cd /usr/g/bin 2. Type sum v*test.py 3. Check that the following output: 19976 23 vdarc_test.py 12195 19 vrac_test.py If this output is viewed, then Class C, p/n 5192235, is correctly loaded If this output is not viewed, then uninstall the current Class C software & load Class C, p/n 5192235 5181591 = 06MW29.7 applications software collector 06MW29.7 software release resolves IG, VIG, DARC2, VDARC computer motherboard End Of Life issue. 06MW29.7 software release supports HS QX/I, LS 1X, 2X, 3X, 4X, 5X, & 7X GRE host consoles, except LS 5X RT16/Xtra
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues