GE Healthcare · CTLightSpeed2600073915010232719
LightSpeed 5.X (80/100kW)
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply. Same error for -5v Analog DAS Power Supply. Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply. Same error # for -5v Analog DAS Power Supply. Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply Wed Feb 22 08:50:01 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 50102/3/0 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply . Measured: 1590 mV, Expected range: 4750 to 5250 mV Tue Aug 22 15:51:09 2006 260007391 4 orp orp_subsys pwrtst.c 90 Tue Aug 22 15:51:08 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 32719/3/0 The Data Acquisition System Has Detected An Error In The -5V Analog DAS Power Supply . Measured: -4748 mV, Expected range: -4751 to -5249 mV
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5126617 ASTEC Analog Power Supply Asm for GDAS 4/8/16 See long description for details 5126617 ASTEC Analog Power Supply Asm for GDAS 4/8/16 Part #: 5126617 Part Type: GP CONSUMABLES Part Weight: 14.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASTEC Analog PS Assembly FRU: Yes Status: Orderable Comments: Added per RT GDAS RSPL, 050210 ms13452
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required