Pular para o conteúdo principal
GE Healthcare · CTLightSpeed182219

Cradle will not drive in or out

Symptom

Cradle will not drive in or out. Cradle will not drive in or out

Error message

Error: 182219 HW error: Home latch driver fault detected by hardware (SMART switch). See LD for Complete Error Fri Jun 3 15:23:41 2005 182219 2 HICT ETC Cradle cmcControl.c 1.23 2/16/99 No Method 296 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:23:42:230 Log Series: 44 HW error: Home latch driver fault detected by hardware (SMART switch).

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2269177 Cradle latch/lock assembly FOR CATHAY-TABLE 2269177 Cradle latch/lock assembly See LD for picture of location 2269177 Cradle latch/lock assembly Part #: 2269177 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: LATCH ASSEMBLY FOR CATHAY-TABLE FRU: Yes Status: Orderable Parts Offered as: Current: 2269177 Used: Not Available Exchange: Not Available Comments: FLS - LATCH ASM ADDED. USED ON H2 TABLE LCC. ASIA HAS IDENTIFIED THIS AS A LINKED PART (SEE 46-296233G2).AWAITING CONFIRMATION ON GL18.MSA8429 11/8/01 START DEPLOYMENT FROM ASIA ON FW7, 02. MS20544 1/31/02
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues