GE Healthcare · Nuclear MedicineInfinia
Infinia table automatic linear motion fails with error 144 axis = 8 Automatic linear motion stops after 90-120…
Infinia table automatic linear motion fails with error 144 axis = 8
Symptom
Infinia table automatic linear motion fails with error 144 axis = 8 Automatic linear motion stops after 90-120cm giving error 144 axis=8. Message on vfd is different if running linear in normal operation (e.g. after greset) or exiting from panel. First case vfd will show error number and message 'gantry failure, perform greset'; in second case you will get error number and message 'velocity test failed' on vfd
Error message
Error #144 (Error 144) Velocity test fail axis = 8 Table Linear
Applies to
- Models
- Infinia
- Product
- Infinia I Infinia I Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software) Infinia II Infinia II Acquisition System
- Component
- *Patient Table *Patient Table *Patient Table
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.