GE Healthcare · CTLightSpeed18554850525118336
LightSpeed 4.X - GOC2
Symptom
CTx00088 Proactive svc call by PDS See long description for details CTx00088 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service
Error message
Error: 185548 The HEMRC operating frequency is below the minimum value. See Long Description for full error Error: 185548 Suite: CT01 Host: OBC Process: Diag Exec File: hemrc_manager.c 1.5 Method: gen_rotor Line: 859 Function: Diagnostics XRAY GENERATION : Tube Rotor Control Test Name: HEMRC Manual Unique ID: 50525 Iteration: 1 Exc. Level: Hard Time: 118336 Error Type: Hardware The HEMRC operating frequency is below the minimum value. Actual frequency: 219.172 Minimum frequency limit: 249.995
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2203602 (replaces 2161307) HEMIT HV TANK used on CT/I (performix), HiSpeed QXi, LS1.X LS2.X, LS 3.X, LS4.X See Long Description for details Part #: 2203602 Part Type: GP REPAIRABLES Part Weight: 55.12 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: HEMIT HV TANK SPARE PART FRU FRU: Yes Status: Orderable Comments: ** FE ASSIST ** The HV adapter/Spacer part number 2171866 Kit that is included with the HV tanks But it is only used on older Highspeed Gantry due to the smaller bolt size. Refer to MI&CT ESERVICE NOTE SN07091114a for additional details. (212001261 21-Jan-2009) *** 2ND OPINION NOTES *** Take an exposure and listen to tank. If you hear an arcing sound most likely bad tank. Rotate Gantry so tank is upside down and take exposure. Listen for arcing. Turn off lights in room and check for a flash from the tank. Remove tank, set on floor on top of the lifting bracket. Make room quiet. Rock tank front to back and side to side, listen carefully. Any sounds, indicates a failure. Check tank for oil leaks Old FRU Compatible No Comments ct/i helio rspl Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable Yes Use on Hands No Use Demand History Yes Ultimate Prime 2203602 New FRU Backward Compatible Yes Old FRU Usable No Change Type Direct ECO Number 9999
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required