GE Healthcare · CTLightSpeed23001900678643520185
91435 LightSpeed 3.X
Symptom
"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors
Error message
Error: 230019006 A non-zero errno was encountered during a remote system call. See Long Description for variable error content Error: 230019006 Host: dhmcvct Process: accs File: sdio_proxy.c Line: 323 A non-zero errno was encountered during a remote system call. The system call may have failed to complete successfully or may have been interrupted. The following diagnostic information was returned. function: read errno: 5 [ Input/output error ] sequence: 2662 Error: 230019006 Host: dhmcvct Process: accs File: sdio_proxy.c Line: 323 A non-zero errno was encountered during a remote system call. The system call may have failed to complete successfully or may have been interrupted. The following diagnostic information was returned. function: mkdir errno: 786435 [ S_ioLib_DEVICE_ERROR ] sequence: 20185 Error: 230019006 Host: dhmcvct Process: accs File: sdio_proxy.c Line: 323 A non-zero errno was encountered during a remote system call. The system call may have failed to complete successfully or may have been interrupted. The following diagnostic information was returned. function: raw_file_allocate errno: 28 [ No space left on device ] sequence: 29 Wed Oct 13 18:25:08 2010 Error: 2300190
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2315116-2 SCSI Cable to HSD used on 2.x, 3.x, 4.x and HiSpeed QXI LightSpeed Systems with a GOC1 or GOC2 console See LD for details 2315116-2 SCSI Cable to HSD used on 2.x, 3.x, 4.x and HiSpeed QXI LightSpeed Systems with a GOC1 or GOC2 consoles SCSI to Drive (Ribbon) 2315116-2 (not identified in manual) For 4.x systems you must also order the adaptor for first time replacement of scsi cable: 2315116-2 (1st time replacement; order adaptor 2376301) (OLD DESIGN 2349539 & 2341109) Part #: 2315116-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: CABLE, SCSI RIBBON - 2 DROP FRU: Yes Status: Orderable Comments: FLS - CABLE, SCSI RIBBON - 2 DROP USED ON CT H16 ADDED PER PCN 270149 MS13452 RISK: SITE DOWN FREQUENCY: HIGH PER PCN 270149 THIS PART IS REPLACEMENT FOR 2349539 PLEASE DO NOT USE ANY STOCK OF2349539 USE ONLY NEW CABLE MS13452 031020 IF FIRST TIME REPLACEMENT PLEASE ALSO ORDER ADAPTOR 2376301 MS13452 CABLE KIT 2333829-2 CONTAINS BOTH THIS CABLE AND ADAPTOR PDR REQUESTED Click here if you need the scsi cable for H3 style console (original style console)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required