GE Healthcare · CTLightSpeedVCT260112011244633260125005
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Gantry shutdown itself for 1 minute and then reset on its own. GANTRY SHUTDOWN ITSELF FOR 1 MINUTE AND THEN RESET ON ITS OWN BETWEEN SCOUTS AND AXIAL SCANS.
Error message
Error: 260112011 Received Table Rx but the Table was NOT READY Error : 260112011 Received Table Rx but the Table was NOT READY Thu Sep 1 08:49:43 2005 Error: 260112011 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr.c Line: 3197 Thu Sep 1 08:49:44 2005 Host : Table Gantry Processor Ermes # : 260112011 Exception Class : Soft Severity : Pri/Soft File : uif_table_tilt_mgr.c Line# : 3197 Function : No System Function Reported Scan Type : Helical Scan: 8439/4/1 Received Table Rx but the Table was NOT READY for Scanning. Error: 244633 Table reports that Table Communication becomes down. Wed May 25 17:35:00 2005 Error: 244633 Host: ct01 Process: fwmgr File: fwmgr_events.c Line: 7301 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Table reports that Table Communication becomes down. Error: 260125005 Table Communication - Watchdog timeout. See long description for complete error Wed Jan 17 08:33:42 2007 Error: 260125005 Host: tgp Process: tgp subsys File: ta callback gt.c Line: 296 Wed Jan 17 08:33:42 2007 Host : Table Gantry Processor Ermes _ : 260125005 Exception Class : Secondary Severity : Pri/Soft File : ta callback gt.c Line_ : 296 Fun
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5120819-2 RoHS Compliant Slipring Brushblock Assembly used on 5.X-7.X gantry. replaces 2349598 5120819 See long description for details 5120819-2 RoHS Compliant Slipring Brushblock Assembly Old FRU Compatible Yes Comments Per ECR 2087298, 212008898, 8/5/09. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5120819-2 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2087298 Part #: 5120819 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT Slipring Brushblock. Same as Hpower, but has additional voltage creepage distance for 208VAC. FRU: Yes Status: Orderable Comments: Added per ECR 2011062 new VCT Brushblock (5120819) on sliprings 2320265 and 2364819. Also added by VCT NPI RSPL UPDATE, ms13452 050916.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.