8-port Ethernet Switch in the System Cabinet.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2249410 A PCI error occured. The PCI Parity Error PERR was asserted. Error: 2249410 File: excLog.c Line: 1132 Task ID: 0x3ac1d68 Time: 23639 Object: AGP A PCI error occured. The PCI Parity Error PERR was asserted. Raven Err Enable Reg: 0x3F; Raven Err Status Reg: 0x88 Raven Err Address Reg: 0x0; Raven Err Attribute Reg: 0x0. Additional information in /usr/g/service/log/vxPciLg-5c2c

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT.
Area
MR

Full cause & step-by-step resolution