Blowers and Fans Not Running in MKS 35KW RF Amplifier Chassis due to the C Phase 8-Amp Fuse Inside MKS 35KW RF Amplifier…
Blowers and Fans Not Running in MKS 35KW RF Amplifier Chassis due to the C Phase 8-Amp Fuse Inside MKS 35KW RF Amplifier Chassis is Blown. 3.0T MKS 35KW RF Amplifier Blowers and Fans Not Running.
Symptom
3.0T MKS 35KW RF Amplifier Blowers and Fans Not Running None of the blowers and fans inside the 3.0T MKS 35KW RF Amplifer are running. All the lights, readouts, and indicators are on but there is zero air flow.
Applies to
- Models
- Excite
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required