GE Healthcare · CTLightSpeed1855533486521538
LightSpeed 4.X - GOC2
Symptom
CTx00088 Proactive svc call by PDS See long description for details CTx00088 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service
Error message
Error: 185553 - Alarm status: 20H (The HEMRC drive has detected a Line Loss. See Long Description for the complete error Tue Aug 10 16:44:06 2004 Error: 185553 Host: OBC Process: Generator Function: XRAY GENERATION : Tube Rotor Control Scan: 34865/2/1 Type: Helical Exception Level: Pri/Most Severe The HEMRC drive has detected a Line Loss. This occurs when the input power to the drive falls below 85% of the nominal Bus voltage. Possible causes: X-Ray tube stator, HEM-IT, HEMRC power supply, .. Alarm status: 20H Or Fri Mar 30 15:21:20 2007 Error: 185553 Function: XRAY GENERATION : Tube Rotor Control Scan: 21538/1/2 Type: Scout Exception Level: Pri/Most Severe The HEMRC drive has detected a Line Loss. This occurs when the input power to the drive falls below 85% of the nominal Bus voltage. Possible causes: - X-Ray tube stator, - HEM-IT, - HEMRC power supply, ... Alarm status: 26H . Note: Alarm Status 26H instead of 20H. This error concept should have additional information for the meaning of the various Hex codes. 20 to 21: Undefined Fault This fault code is undefined and may indicate a defective Drive. Retry operation. If problems persist contact the factory for instructions. or 20:
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2179860-2 HEMRC CONTROL BOARD See LD for more info 2179860-2 HEMRC CONTROL BOARD. Part #: 2179860-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HEMRC CONTROL BOARD FRU: Yes Status: Orderable FEMC PART COMMENTS: FLS - HEMRC CONTROL BOARD. REPLACES P/N 2179860. ADDED PER PCN 220717.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required