GE Healthcare · CTLightSpeed26000740250444260007774
LightSpeed 4.X - GOC2
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260007402 Converter Board(s) Asserted a Board FAULT. See Long Description for complete error Tue Dec 28 09:41:31 2004 Error: 260007402 Host: orp Process: orp_subsys File: cnvmgr.c Line: 605 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260007402 Exception Class : Abort Severity : Secondary File : cnvmgr.c Line# : 605 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 Converter Board(s) Asserted a Board FAULT. Error: 260007774 List of Converter Cards From the Previous Error Message Card List: xx xx See Long Description for full error Tue Dec 30 09:25:17 2008 Error: 260007774 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 852 Tue Dec 30 09:25:17 2008 Host : Das Ermes # : 260007774 Exception Class : Secondary Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 852 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 List of Converter Cards From the Previous Error Message Card List: 8,11,14-15,19,21,25-27,29-36,38-58,60-63 Note on this error. Sometimes the actual bad board will be the one that is not listed.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2355552 MDAS 16 Convertor Card ( replaces 2322994-2 ) Used In LightSpeed 16 Gantry with MDAS. 2355552 MDAS 16 Converter Card used in LS 4X with 16 slice MDAS (replaces 2322994-2) Part #: 2355552 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.25 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: SERVICE, MDAS16 CONV BD FRU: Yes Status: Orderable Comments: Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money. This is an MDAS16 CONV BD .compared to the SDAS. Make sure the correct Convertor is Ordered and replaced.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required