GE Healthcare · MRSigna2225660

Erbtec Faults Due to Door Knob Capacitor Arcing In RF Amplifier.

Symptom

Unable to scan with Erbtec errors

Error message

2225660 ERBTEC RF Amplifier Fault 81 3kV Arc Detected Thu Oct 23 19:32:24 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225660 File: PMH.m 1.29 Method: spi_rf_mon Line: 3035 ERBTEC RF Amplifier fault 81. 3kV arc detected. Possible defective IPA tube or bad processor board.

Applies to

Models
Signa
Product
Signa Horizon (5X Horizon) 5X Horizon systems
Software
5.X
Area
MR