Pular para o conteúdo principal
GE Healthcare · CTLightSpeed182004502384311094308571821251598546

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

No flashing eights on gantry display and Scanning hardware reset failed

Error message

Error: 182004 LAN Watchdog TIMEOUT See LD for the full error code Mon Feb 14 10:28:30 2005 Error: 182004 Suite: SJ01 Host: ETC Process: Local Exc Handler File: exl_lan_in_mail.c Method: No Method Line: 175 Function: No System Function Reported Scan: 50238/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:28:30:550 Log Series: 31 LAN Watchdog TIMEOUT. HW/SW Error: Did not receive LAN Watchdog message from STC. (Current Time:431109 Last Wdog LAN Probe Time:430857 (in Ticks))

Error: 182125 LAN Watchdog TIMEOUT with OBC or ETC (uuid 93, 193) (SW opening abort relay). See LD for full errors Mon Dec 6 22:19:33 2004 Error: 182125 Suite: CT01 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Method: No Method Line: 773 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 22:19:32:730 Log Series: 27 LAN Watchdog TIMEOUT with OBC (uuid 193) (SW opening abort relay). HW/SW Error: Did not receive LAN Watchdog message from OBC. (Last Wdog Probe Sent:1598546 Last Response Rcvd:1598350 (in Ticks)). Error could be for the ETC also Reference Thu Jan 12 09:54:50 2006 Error: 182125 Host: STC Process: Master Exc Handler Fi

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2224296-3 Artysen board for 1X/2X/3X/4X used in the ETC/STC/OBC See Long Description for FEMC information and board locations Part #: 2224296-3 Part Type: GP REPAIRABLES Part Weight: 1.09 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: VE-GE, 8MB, Artesyn CPU Bd FRU: Yes Status: Orderable Parts Offered as: Current : 2224296-3 Harvested : Not Available Exchange : Not Available Comments: ** FE ASSIST ** Repair operations has found that 68% of the time the Artesyn board (2224296-3) is returned for repair, it has No Defect Found. Check switch settings (212001261 2/12/09)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues