GE Healthcare · MRExciteSigna2244655
Bright line (horizontal or vertical) through image due to gradient switch.
Symptom
MRIQ Bright Line Artifact and GSC/Gradient Errors. See attachment .jpg file. Link to image quality troubleshooting guide. http://3.87.118.28/optec3/common/8xdocs/epst/mod11/m11_010_000.html
Error message
2244655 Gradient Switch Fault Detected. Scan will be paused 2244655 Host: TwinExciteLXMR Process: NSP File: SCP:Grad Line: 704 Gradient Switch Fault Detected. Scan will be paused.-> Extended Error: Please restart the scan (continuing the scan might cause image quality issues). If the fault persists,replace the Gradient Switch
Applies to
- Models
- Excite, Signa
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- 2253466 TWIN GRADIENT SWITCH Module Box Same part number for 1.5T and 3.0T. Same part number for 1.5T and 3.0T.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.