GE Healthcare · CTLightSpeed2600024612600010235072326000235850011260003117
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260002461 Data Parity Error detected See long description for full error Sun Oct 7 06:29:24 2007 Error: 260002461 Host: DAS Process: DCB Firmware File: parity.c Line: 228 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 7758/3/1 Type: Helical Exception Level: Sec/Soft Time: 06:29:24:675 Log Series: 165 SCAN ABORT -- DAS Has Detected An Error.^B Data Parity Error detected in the 2A ODD Data Stream(s) on Converter Board 48 or the DCB Board. (CSI:1 RSM:0x2). Do Not Use - use one of the other two Error: 260001023 See Long Description for complete error NOTE: The error message code included in this error is variable. You will need to refer to the log to find the cause of this error. Tue Dec 7 07:23:22 2004 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 367 Function: No System Function Reported Scan: 50723/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 07:23:22:910 Log Series: 37 The DCB has notified the system of an error condition which prevents scanning. See log for further information: Error message code: 260002358 Notification level: 5 Sequence number: 4 Fri May 11 06:44:04 2007 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line:
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2137958-4 SDAS Convertor Board (replaces 2137958-3) May use 2137958-3U converter Board if no current stock of the 2137958-4 See long description for details Part #: 2137958-4 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: -39999.60 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: CONVERTOR WITH BLACK BEUTY CONNECTOR FRU: Yes Status: Orderable Comments: OBSOLETE PART case 10788898 This part has been identified as obsolete and no short-term solution has been identified FE must identify and order a higher or lower level spare part as required Contact your support engineer on-line engineer or Parts ID team should you require support 212012173 06-26-07 REPLACES P/N 2137958-3 THE NEW -4 IS COMPATIBLE TO THE -3. MS20544 ON 3/7/03. ** FE ASSIST ** May use the USED part 2137958-3U converter Board if no current stock of the 2137958-4. (9/10/2007 212001261)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required