Scanner fails to boot to applications and is Unable to verify DARC system startup

Symptom

Scanner fails to boot to applications, Unable to verify DARC system startup. The system will not boot up to applications with a pink attention box: Unable to verify system functionality on startup. Please power cycle the console. If the system fails to startup after the power cycle, then please call GE Service. FE or OLE sees that they can telnet to the DARC but they cannot ping or rsh with the DARC. If the DARC is observed while booting, Drive X adapter X predicting future failure Press any key to continue is seen.

Error message

Do Not Use - use one of the other Error: 290014004 Unable to verify DARC system startup See Long description for complete error NOTE: This is error does not mean you need to replace the darc. Troubleshooting is necessary This error comes in different contexts per the below Mon Dec 12 08:00:11 2005 Error: 290014004 Host: ct99 Process: gre-verify-darc File: gre-verify-darc.cpp Line: 172 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - waiting for ping darc response (exit status: 8) Fri Dec 8 14:42:12 2006 Error: 290014004 Host: kghcateqxi Process: gre-verify-darc File: gre-verify-darc.cpp Line: 179 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - scan disk array is non-recoverable (exit status: 22) Fri Mar 16 09:13:03 2007 Error: 290014004 Host: ctflsh Process: gre-verify-darc File: gre-verify-darc.cpp Line: 179 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - waiting for rsh darc response (exit status: 9) Fri Mar 14 05:49:37 2008 Error: 290014004 Host: nvgect01 Process: gre-verify-darc File: gre-verify-darc.cpp Line: 180 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - waiting

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2362873 SDDA (Scan data disk array) used on GOC3. See Long Description for details. The GOC3 to GOC4 upgrade kits are now under 1 part number, 5144058-500 plus the FE will have to also order the DARC2 that is needed by the DAS as a separate order. Part #: 2362873 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: SCAN DATA DISK ARRAY - GRE FRU: Yes Status: Orderable Comments: OBSOLETE PART case 8391074 This part has been identified as obsolete and no short-term solution has been identified FE must identify and order a higher or lower level spare part as required Contact your support engineer on-line engineer or Parts ID team should you require support 212012173 05-23-07 ** FE ASSIST ** For CT not PET.. IF no stock of 2362873 SDDA must upgrade to DARC II + ICOM Kit per one of the following part number's * 5144058-500, 16 MDAS 16,8,4 GDAS DARC Replacement kit, JARRELL * 5144058-501, 4,8 MDAS DARC Replacement Kit, JARRELL * 5144058-502, 4 SDAS DARC Replacement Kit, JARRELL Documentaion 5194645-100 Must order software release 06MW29.7, p/n 5193989 and perform a LFC to support this FRU if system is not currently running 06MW29.7 or better (2/19/07 212001261) *** 2ND OPINION NOTES *** Look at the DARC messages log for SCSI disk errors located at /var/log/messages
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.