Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230019002

89623 LightSpeed 3.X

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230019002 The scan disk I/O layer failed to consume a response from the scan disk server. See LD for complete error Tue Nov 2 03:19:30 2004 Error: 230019002 Host: sgredct1 Process:scanfilemgr File: sdio proxy.c Line:174 The scan disk I/O layer failed to consume a response from the scan disk server. function:consume sequence:6708

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2315116-2 SCSI Cable to HSD used on 2.x, 3.x, 4.x and HiSpeed QXI LightSpeed Systems with a GOC1 or GOC2 console See LD for details 2315116-2 SCSI Cable to HSD used on 2.x, 3.x, 4.x and HiSpeed QXI LightSpeed Systems with a GOC1 or GOC2 consoles SCSI to Drive (Ribbon) 2315116-2 (not identified in manual) For 4.x systems you must also order the adaptor for first time replacement of scsi cable: 2315116-2 (1st time replacement; order adaptor 2376301) (OLD DESIGN 2349539 & 2341109) Part #: 2315116-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: CABLE, SCSI RIBBON - 2 DROP FRU: Yes Status: Orderable Comments: FLS - CABLE, SCSI RIBBON - 2 DROP USED ON CT H16 ADDED PER PCN 270149 MS13452 RISK: SITE DOWN FREQUENCY: HIGH PER PCN 270149 THIS PART IS REPLACEMENT FOR 2349539 PLEASE DO NOT USE ANY STOCK OF2349539 USE ONLY NEW CABLE MS13452 031020 IF FIRST TIME REPLACEMENT PLEASE ALSO ORDER ADAPTOR 2376301 MS13452 CABLE KIT 2333829-2 CONTAINS BOTH THIS CABLE AND ADAPTOR PDR REQUESTED Click here if you need the scsi cable for H3 style console (original style console)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.