GE Healthcare · CTHiSpeed

Octane would not boot with PANIC

Symptom

System would not boot with PANIC: KERNEL FAULT reported. System would not boot and monitor screens turn blank with PANIC: KERNEL FAULT.

Error message

Memory Parity Error in SIMM Sxx Recoverable memory parity error detected by CPU PANIC: KERNEL FAULT - found in SYSLOG See LD for more info Memory Parity Error in SIMM S10 Recoverable memory parity error detected by CPU PANIC: KERNEL FAULT Defective SIMM memory on CPU board caused Memory Parity Error in the SYSLOG Troubleshooting kaiserct 12# cd /usr/adm kaiserct 13# tail -100 SYSLOG | more Jul 11 12:35:05 3D:kaiserct ppp[308]: + killall -v -USR1 gated Jul 11 12:35:05 3D:kaiserct ppp[308]: + 1>& 2 Jul 11 12:35:05 3D:kaiserct ppp[308]: UX:killall: ERROR: process <gated> not found Jul 11 12:35:06 2E:kaiserct savecore: reboot after panic: <0> PANIC: KERNEL FAULT Jul 11 12:35:06 2E:kaiserct savecore: pb 0: 8 <0x380> code:30 Jul 11 12:35:06 2E:kaiserct savecore: pb 1: Memory Parity Error in SIMM S10 Jul 11 12:35:06 2E:kaiserct savecore: pb 2: Recoverable memory parity error detected by CPU Jul 11 12:35:06 2E:kaiserct savecore: pb 3: Memory Parity Error in SIMM S10 TYPE THE FOLLOWING FOR SUSPECTED “MEMORY ERRORS ONLY” IN THE SYSLOG kaiserct 12# cd /usr/adm kaiserct 11# cat SYSLOG | grep Memory Jul 11 12:12:17 2A:kaiserct unix: Memory Parity Error in SIMM S10 Jul 11 12:13:26 2A:kaiserct un

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2115457-22 16MB SIMM or 2115457-23 32MB SIMM See LD for Details 2115457-22 16MB SIMM or 2115457-23 32MB SIMM See Picture Below: Part #: 2115457-22 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 16MB SIMM MEMORY MODULE FRU: Yes Status: Orderable Comments: FLS - 16MB SIMM MODULE; ALTRON CPU CARRIER RHAPSODE. CT EXPECTS SPIKE IN DEMAND 1ST QTR 2001, PER DAVID MAPES. MS7093 HQ: 1 OF 4 16MB SIMMS. FRU2 64MB SIMM 2115457-9 **THIS PART ONLY HAS 1 SIMM** 16 MB IF NEED A SET ORDER 4 GELS - 16MB SIMM MODULE; ALTRON CPU CARRIER RHAPSODE. C1-&gt;C3 10-98 SPS: CQA 985680 CAUSED STOCK PURGE. SEE -------------------&gt; D915 HQ: 1 OF 4 16MB SIMMS. FRU2 64MB SIMM 2115457-9 **THIS PART ONLY HAS 1 SIMM** 16MB IF NEED A SET ORDER 4,16MB SIMM MODULE Part #: 2115457-23 Part Type: GP CONSUMABLES Part Weight: 0.10 LBR Repairable: No Returnable: No Hazardous Material: N Description: 32MB SIMM MEMORY MODULE FRU: Yes Status: Orderable Comments: FLS - 32MB SIMM MODULE 1 SIMM ALTRON CPU CARRIER RHAPSODE REPLACED BY 2149503 PER KENT JOHNSON 3/97 MS08992 GVR HQ:1 OF 4 SIMM MODS IN FRU 2, 2115457-10 PCN 189977 FLS - DEFECTIVE PARTS MUST BE RETURNED TO: GEORGE FARRINGTON 16800 WEST RYERSON ROAD NEW BERLIN, WI 53151 THIS IS PART OF A BLACK BELT PROJECT!!
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.