Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000777127259

Intermittent DAS communication errors.

Symptom

Intermittent DAS communication errors. Problem occurs if scanning is paused or aborted. See Long Description Once fault occurs the exam needs to be ended and a scan hardware reset performed at a minimum to resume scanning again.

Error message

Error: 260007771 DAS Single Card Communication Error See LD for complete error Error: 260007771 DAS Single Card Communication Error (could be MDAS or GDAS) MDAS: Mon Apr 17 08:16:46 2006 Error: 260007771 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 863 Mon Apr 17 08:16:45 2006 Host : Das Ermes # : 260007771 Exception Class : Soft Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 863 Function : No System Function Reported Scan Type : None/Unknown Scan: 1094/2/0 M-DAS Single Card Communication Error. See Following Error Message for Converter Card List Board:19 Command: State Query Status: Communication timeout - no response. GDAS: Sat Aug 16 14:08:17 2008 Error: 260007771 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 863 Sat Aug 16 14:08:16 2008 Host : Das Ermes # : 260007771 Exception Class : Secondary Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 863 Function : No System Function Reported Scan Type : Scout Scan: 27259/1/1 Converter Board Single Card Communication Error.^C See Following Error Message for Converter Card List Board: 81 Command: State Query Status: Communication timeout - no response.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues