GE Healthcare · CTHiSpeedLightSpeedBrightSpeed100331019116896
HiSpeed CTi 6.X with Octane and HSA Gantry
Symptom
Corrupt media appears in message box: Cannot attach MOD for archiving.
Error message
Error: 10033 There is no media in the drive. See Long Description for complete error Sun Aug 14 11:40:26 2005 Error: 10033 Host: ct01 Process:arsjob File: ArchiveDevice.c Line:4781 There is no media in the drive. Error: 10191 Failed save exam. Mon Nov 27 03:45:02 2006 Error: 10191 Host: emc1 Process: arserver File: arscommand.c Line: 2304 Failed save exam. Paused 16896
Applies to
- Models
- HiSpeed, LightSpeed, BrightSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2260734-2 MOD drive Used In cti 6.x octane console 2260734-2 MOD drive Used In cti 6.x octane console MAXOPTIC MOD DRIVE, P/N 2260734-2. SAME AS T3202KP. BOTH REPLACED BY SONY MOD DRIVE, P/N 2318598 PER PCN 220399 ON 12/15/02. MS20544 MAGNETO OPTICAL DRIVE 4.2 FIRMWARE VERSION, USED ON HELIOS. REPLACED 2260734, PER PCN 199368 IF YOU HAVE INDIGO II SYSTEM ORDER 2260734-2. IF YOU HAVE QX/I SYSTEM, MUST INSTALL FMI 25261 1.1 SW. IF YOU HAVE OCTANE CTI, ORDER 2268089 KIT.THE 2268089 KIT CONTAINS 5.5 SW AND THE STAR MAXOPTIC DRIVE 4/30/01. REPLACED BY 2280678 PER PCN 220117. FIRMWARE UPGRADE FROM 4.2 TO 4.5. 2260734-2 (T3202KP). REPLACED BY SONY MOD DRIVE PER PCN 220399. 12/15/02. John Kreczmer 8/24/04
- Area
- CT
Full cause & step-by-step resolution