Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260107214

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260107214 Gantry Failed to Stop at Position XX degrees within XX.X seconds See Long Description for complete error Thu Sep 22 13:19:47 2005 Error: 260107214 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3687 Thu Sep 22 13:19:47 2005 Host : Table Gantry Processor Ermes # : 260107214 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 3687 Function : AXIAL CONTROL : Axial Control Scan Type : Scout Scan: 279/1/1 Gantry Failed to Stop at Position 90 degrees within 28.000 seconds. (AMC state:7 Cmd:24 CurPos:91.0) Retry Rx to verify probable Axial Drive or Gantry Balance or Brake or Encoder Hardware.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.