GE Healthcare · MRSigna22540512254052

LX System has power monitor trips on virtually every scan.

Symptom

Unable to scan with Power Monitor Faults

Error message

2254051 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. See Long Description Fri Aug 20 08:33:11 2004 Error: 2254051 Host: lx-mr Process: NSP File: SCP:PmMdsHWO Line: 1019 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 35 pulse width: 33 duty cycle: 09 Fault in head amplitude. This could also occur when power cycling SSM.

2254052 Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and Patient Weight Fri Aug 20 08:54:29 2004 Error: 2254052 Host: lx-mr Process: NSP File: SCP:PmMdsHWO Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 35 pulse width: 33 duty cycle: 09 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR