GE Healthcare · CTLightSpeed260007391
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Error: 260007391The Data Acquisition System Has Detected An Error In The Analog Ground 2 DAS Power Supply. Wed Apr 5 10:58:29 2006 Error: 260007391 Host: orp Process: orp_subsys File: pwrtst.c Line: 90 Wed Apr 5 10:58:28 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The Data Acquisition System Has Detected An Error In The Analog Ground 2 DAS Power Supply. Measured: 686 mV, Expected range: -100 to 100 mV
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2289460 MDAS CHASSIS, RIGHT ASM -H3 See Long Description for Additional information. Part #: 2289460 Part Type: GP CONSUMABLES Part Weight: 7.52 LBR Repairable: No Returnable: No Hazardous Material: N Description: BACKPLANE TO CHASSIS MDAS RIGHT FRU: Yes Illustration 2: DAS Chassis Mounting Bolts (Between Left and Center Chassis) The center chassis shares one bolt each with the left and right chassis. When removing the left or right chassis, the second bolt on the shared side of the center chassis must be loosened slightly. Be careful not to damage detector flexes.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required