GE Healthcare · CTLightSpeedVCT260134802398728100000

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Cradle / IMS on GT1700 / GT2000 would not advance from home and makes grinding noise. See Long Description for more details and sound. Cradle / IMS on GT1700 / GT2000 would not advance from home and makes grinding noise This problem happen after moving a large patient off of the table, or after physically moving the IMS to it’s mechanical out of gantry limit (home). Attachment is the sound the table makes when the cradle move button is pushed. The button was pressed 2 times (2 sounds)

Error message

Error: 260134802 IMS position is out of tolerance. See Long Description error messages Thu Oct 20 08:01:13 2005 Error: 260134802 Host: tgp Process: tgp subsys File: gtcb movechk.cxx Line: 581 Thu Oct 20 08:01:13 2005 Host : Table Sub-System Ermes # : 260134802 Exception Class : Secondary Severity : Pri/Soft File : gtcb movechk.cxx Line# : 581 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 IMS position is out of tolerance. current position(um) : 398728 target position (um) : 0 start position (um) : 398728 velocity (um/s): 100000 time from start (ms) : 300

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5168472-2 New style IMS bearing replacement kit for GT1700 or GT2000 tables. See long description for details Part #: 5168472-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: IMS DRIVE REPLACEMENT SET Positioning GT FRU: Yes Status: Orderable There is a new Kit for the VCT Table IMS. The old kit is no longer available. The old Kit part number will automatically change up to the number for the new kit: 5168472-2 . FE's are confused by the parts in the Kit and there are no installation instructions. FMI 26809 will be released with instructions: The Kit contains only one Bearing, but the Table has 2 Bearings. This is correct . The new Bearing is larger and as a result only one Bearing is required now. There are no instructions in the Kit on how to install all the parts. Use the hyperlink below to get instructions on how to install the Kit. http://jptyoiis01.ap.health.ge.com:8000/ctdoc/FMI/19/26809.pdf
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues