Pular para o conteúdo principal
GE Healthcare · MRExcite222381722238592223854

HDx TPS Reset Fails Due to Cat-6 Ethernet Cable from ICN and Ethernet Switch.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2223859 The TPS Reset Failed because of a VRE Reset Failure See long description. 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Extended Error: This indicates that a failure to reset the VRE properly caused the TPS Reset to fail. Please refer to the VRE Troubleshooting Guide for further troubleshooting this issue, which can be found in the MR Service Documentation.

2223854 Extended TPS Reset in progress (up to 8 minutes). Please be patient.-> See Long Description 2223854 ERROR: EM_mrNSP_VRE_RESET_PROLONGED Extended TPS Reset in progress (up to 8 minutes). Please be patient. Extended Error: One or more ICNs of the VRE might have been non-responsive during a TPS Reset. We therefore reboot all ICNs in an attempt to recover. The operator will experience longer TPS Reset times because of this.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Component
5160775 FRU ICN Ethernet Cable Collector This kit contains: 3 cables between ICN and Ethernet Switch, 1 J63 to Ethernet Switch cable, 1 Ethernet cable 100 ft (30.5m) in length between Host and J63 on System Cabinet, and 1 cable label book. HDx system.
Area
MR

Full cause & step-by-step resolution