U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2254202 SRFD2 RF Amplifier Internal Cable Fault Error: 2254202 Host: LX MR1 Process: NSP File: HPC PORT ID:spi rf mon Line: 1717 SRFD2 RF Amplifier Internal Cable Fault. The RF Amplifier is reporting a cable is disconnected internally. Check internal cabling behind front panel of SRFD2. If checks okay, then replace RF Amplifier.

2254216 SRFD2 RF Amplifier Junction Temp Fault.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
Area
MR

Full cause & step-by-step resolution