Pular para o conteúdo principal
GE Healthcare · CT178575178517244518244555

Lightspeed 1.X - Octane

Symptom

NO flashing eights when hardware reset attempted. NO flashing eights when hardware reset attempted

Error message

Error: 178575 Failed to R-HARD the following controllers: STC OBC ETC. See long description for full error Fri Oct 8 07:49:44 2004 178575 2 ct03 fwmgr HwSetUp.c 837 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to R-HARD the following controllers: STC OBC ETC.

Error: 178517 Failure writing to socket connected to controller STC/OBCR/ETC. No route to host See Long Description for variations Fri Oct 8 07:49:17 2004 178517 ct03 fwmgr NBSSocket.c 690 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failure writing to socket connected to controller STC. OS error : 113 --- No route to host Fri Oct 8 07:49:13 2004 178517 ct03 fwmgr NBSSocket.c 690 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failure writing to socket connected to controller OBCR. OS error : 113 --- No route to host Fri Oct 8 07:49:33 2004 178517 4 ct03 fwmgr NBSSocket.c 690 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failure writing to socket connected to controller ETC. OS error : 113 --- No route to host

Error: 244518 Scan hardware reset faile

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2377227 UTP Transceiver or Media Convertor used on OC with 1X - 4X Gantry - replaces 2148559 10BASE-2 TO UTP CONVERTER See Long Description for Details. 2377227 UTP Transceiver or Media Covnerter for GRE GOC3 Consoles with HP-8000 CPU and older style Gantries LS1X - 4X Part #: 2377227 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: UTP TRANSCEIVER - BNC TO RJ45 With I.T.E PWR SPPLY 24000027 FRU: Yes Status: Orderable Part #: 2148559 Part Type: GP CONSUMABLES Part Weight: 1.36 LBR Repairable: No Returnable: No Hazardous Material: N Description: 10BASE-2 TO UTP CONVERTER FRU: Yes Status: Orderable Comments: ***OBSOLETE PART Case # 11345706*** This part has been identified as technically obsolete and no short term solution has been identified. FE must identify and order a higher or lower level spare as required. Contact your support engineering, On-line Engineering or Parts ID team should you require support. 212012173 7-24-49 ** FE ASSIST ** For P/N 2148559 - Please order the newer media converter 2377227 and external terminator 2154019. The 2154019 50 ohm external terminator is required.(212001261 10/2/07) I re addded my note 7/28/08 as someone removed it
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required