Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000680450004260006806

LightSpeed 4.X - GRE

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260006804 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration See long description for complete error Thu Nov 18 20:34:37 2004 Error: 260006804 Host: STC Process: Axial File: amc_exec.c Line: 3606 Function: AXIAL CONTROL : axial communications Scan: 50004/1/0 Type: None/Unknown Exception Level: Secondary Time: 20:34:37:420 Log Series: 37 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry.

Error: 260006806 Axial Motor Control (AMC) logic will NOT accept commands without a Valid Configuration/Characterization file. See the Long Description for the full error code Thu Nov 18 20:34:37 2004 Error: 260006806 Host: STC Process: Axial File: amc_exec.c Line: 3464 Function: No System Function Reported Scan: 50004/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 20:34:37:390 Log Series: 37 Axial Motor Control (AMC) logic will NOT accept commands without a Valid Configuration/Characterization file. (Cmd: 0x1804ffff).

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P15 (250V,8A) Fuse used in axial dynamic brake circuit as CT2A1A10F1 and found in FAN ASSY- GANTRY, and F4, F5 on HEMRC Interface Board and other places See long description for details Part #: 46-170021P15 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 008.000A 0250V 3AB SLOW UL/ FRU: Yes Status: Orderable Comments: FLS - FUSE 8 AMP - 250 VOLT 3AB
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues