Pular para o conteúdo principal
GE Healthcare · CTLightSpeed290009018290009015

87109 LightSpeed 4.X

Symptom

A Pink Pop up message is reported on boot-up that Recon times may be slower - scanning may continue See Long description to see the Pink Pop up message A Pink Pop up message is reported on boot-up that Recon times may be slower

Error message

Error: 290009018 All the IG nodes are not operational. See long description for full error Mon Nov 29 06:50:22 2004 Error: 290009018 Host: darc Process: recon control File: recon control cb.cxx Line:1549 All the IG nodes are not operational

Error: 290009015 Recon connect timer triggered on 3 ( could also be 1 or 2 ) See LD for full error Error: 290009015 Recon connect timer triggered on 3 Tue Aug 22 07:26:18 2006 Error: 290009015 Host: darc Process: recon_control File: recon_control_cb.cxx Line: 1605 Recon connect timer triggered on X X = the number of ig that failed ( i.e. 1 , 2 or 3)

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.