Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230015037200108011200108010

LightSpeed 3.X (Ultra, Plus, QX/i) -Octane

Symptom

Fastcal scan aborts. Problem finding a calibration vector or calibration file.

Error message

Error:230015037 /usr/g/cal_db/kv120/large_sfov/1x125/large_spot/dasgain8/cal.1.acal has size: 0 See Long Description for complete error Mon Jun 20 07:06:45 2005 Error: 230015037 Host: ctxm Process: accs File: cal_db_utils.c Line: 1078 /usr/g/cal_db/kv120/large_sfov/1x125/large_spot/dasgain8/cal.1.acal has size: 0

Error: 200108011 The following calibration vector is not valid: See Long Description for complete error Mon Jun 20 07:06:46 2005 Error: 200108011 Host: ctxm Process: scanRx File: CalStatusInfo.cxx Line: 554 Function: Data Acquisition : OC Processing The following calibration vector is not valid: kV:120 sfov:SFOV_LARGE collimation:0 spotsize:LARGE_SPOT.

Error: 200108010 This system has not been fully calibrated for XXX kV. You will not be able to scan at XXX kV until the system has been calibrated for this kV. See Long Description for complete error Mon Jun 20 07:06:46 2005 Error: 200108010 Host: ctxm Process: scanRx File: SIRx.cxx Line: 2380 Function: Data Acquisition : OC Processing This system has not been fully calibrated for 120 kV. You will not be able to scan at 120 kV until the system has been calibrated for this kV.

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required