Auto Prescan failed, RF Amplifier is not ready.

Symptom

Auto Prescan failed, RF Amplifier is not ready Getting another message that RF Amplifier is powering but message never goes away.

Error message

2224666 Auto Prescan Failed, RF Amplifier is Not Ready NOTE: When using this error to create new knowledge, please include any associated error codes. 2224666 is a generic error code which requires further information. Mon May 17 08:19:08 2004 Error: 2224666 Host: lx_mr Process: psc File: PscMainControl.cc Line: 1296 Auto Prescan failed, RF amplifier is not ready.

2254250 - 3T RF Amplifier 50 Volt Supply Low (fault 35). The SSD 50 Volt supply was below 5% specification. See Long Description for extended error message. 2254250 3T RF Amplifier 50 Volt Supply Low (fault 35). The SSD 50 Volt supply was below 5% specification. Possible Causes: Defective supply or short on supply output. If the fault persists, replace the RF amplifier. Extended Error: The RF Amp 50 V Power Supply has low voltage. Check fuses F15, F16 and F17, replace if necessary. If the fuses are functional, replace the power supply deck.

Applies to

Models
Excite
Product
3T Excite G3 Software ONLY
Software
E2 3T Eclipse
Component
5125603-2 E2 3T RF Amp Power Deck FRU E2 3T RF Amp Power Deck FRU 5125603-2 E2 3T RF Amp Power Deck FRU 5125603-2