GE Healthcare · MRExcite2225684

The Neslab gradient coil chiller was not running.

Symptom

Will Not Scan With Patient Comfort Sensor Trip. See Long Description Patient comfort sensor trip; cannot scan due to bore sensor indicating temperature is too hot for patient safety. A message "Patient Comfort Sensor Trip" appears in the message box on the left side of the scan page at the start of autoprescan.

Error message

2225684 Scan inhibited. Patient comfort sensor trip. See Long Description 2225684 ERROR: EM_SPI_BORE_TEMP_TRIP_2 Scan inhibited. Patient comfort sensor trip. To return to scanning, verify: 1. Patient fan is ON. 2. Scan room temperature is normal. 3. Air flow through the bore is unobstructed. 4. Gradient Cooling fan is normal. 5. BRM heat exchanger is normal.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR