Pular para o conteúdo principal
GE Healthcare · MRExcite2218592

Fails TPS Reset when booting up to Applications or when Resetting TPS.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2218592 The MrMail send of reset to TPS/AGP timed out Thu Oct 4 07:04:55 2007 Error: 2218592 Host: rmnt Process: scn File: msaUtil.m Line: 175 The MrMail send of reset to TPS/AGP timed out.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT.