GE Healthcare · MRExciteDiscoveryOptima22487592224666
CWA ID86581. Will Not Scan With RF Errors.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2248759 RF Amplifier Response Timeout RF MDS peripheral to RF Amp/RF Interface link failure. 2248759 NSP SCP: RF amplifier response timeout. RF MDS peripheral - RF Amp/RF Interface link failure. Verify that the RF Amplifier or RF Interface is powered on and that the RF MDS peripheral is connected to the RF Amplifier or RF Interface. 2224666 Auto Prescan Failed, RF Amplifier is Not Ready NOTE: When using this error to create new knowledge, please include any associated error codes. 2224666 is a generic error code which requires further information. Mon May 17 08:19:08 2004 Error: 2224666 Host: lx_mr Process: psc File: PscMainControl.cc Line: 1296 Auto Prescan failed, RF amplifier is not ready.
Applies to
- Models
- Excite, Discovery, Optima
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.