GE Healthcare · MRExcite2225935

Defective Fiber Optic Cable Run # 712.

Symptom

Gating does not work and the message The PAC does not respond displayed during system boot or TPS reset. This problem was found while installing this new system. Ran Diagnostics --> Peripherals --> PAC Diagnostics --> PAC Power up Test and SPU-->PAC Loop back Test, both of which failed with error codes 2254702 & 2254703. Measured PAC power at rear of SSM module and found it to be okay. This problem was found while installing this new system. Ran Diagnostics --> Peripherals --> PAC Diagnostics --> PAC Powerup Test and SPU-->PAC Loopback Test, both of which failed with error codes 2254702 & 2254703. Measured PAC power at rear of SSM module and found it to be okay.

Error message

2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. 2225935 The PAC does not respond. Check PAC power and fiber optic connections, run PAC diags. This error condition adversely affects gated scans only. 2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. This error condition adversely affects gated scans only.

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR

Full cause & step-by-step resolution