System won't scan with Loop Contactor error

Symptom

System won't scan with Loop Contactor error 183949 and 183921 Tech had completed the scout and had begun the exam when she started getting resumes.She ended the and restarted the exam but now could not scout. Error Log: Axial Error: 183949 File: ax error.c Method: No Method Line: 1106 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Scan: 4718/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 10:41:19:725 Log Series: 673 Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor. Axial Gantry Command/Status:0x9e08. Wed Apr 21 10:38:43 1999 Suite: STJ1 Host: STC Proc: Axial Error: 183921 File: ax error.c Method: No Method Line: 1106 Function: AXIAL CONTROL : Axial Loop Contactor interlocks Scan: 4717/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:38:39:730 Log Series: 659 Axial conditions at FAULT: velocity fb = -1 position fb = -55075 Formatted data: 8100 100 10 1 1 0 1 0 0 1 1 1064 346112 55075 18 -19 127 128 0x009e28

Error message

Error: 183949 Could NOT CLOSE Gantry (Axial) Loop Contactor See LD for full error CT01 Host: STC Proc: Axial Error: 183949 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor Another example: Tue May 24 08:41:21 2005 Suite: CT02 Host: STC Proc Axial Error: 183949 File: ax_error.c Method No Method 1114 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Scan: 65001/11/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:21:940 Log Series: 60 Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor. Axial Gantry Command/Status:0x9e28. EN 401 SR 442

Error: 183921 Axial conditions at FAULT: See Long Description for complete error Mon Oct 17 08:26:48 2005 Error: 183921 Suite: SMCT Host: STC Process: Axial File: ax_error.c Method: No Method Line: 1114 Function: AXIAL CONTROL : Axial Servo Control Loop Scan: 28848/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 08:26:47:775 Log Series: 141 Axial conditions at FAULT: velocity_fb = 2662 position_fb = 50389 Formatted data: 8100 100 8 7 1 0 1 1 101 1 0 2662 -12655275 13081394 2741 2582 134 133 0x019e3

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-296690P1 auxillary contact sense switch on the side of PDU K2 contactor 46-186852P2 See long description for details Part #: 46-296690P1 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: PIGGY SWITCH K2 RELAY FRU: Yes Status: Orderable Comments: FLS - N.O. AUXILLIARY CONTACT (INTERLOCK) USED ON ALL HS PDU'S FRU CHANGE PER DUANE D. 10/3/97 PCN TO FOLLOW 2WENN 10/3/97 MS08992 PCN 199098 MS4636 10/20/97 ************************************************************CPD USE ONLY: IF NO PARTS AVAILABLE, BROKEN SWITCH CAN BE REPAIRED. FE # 1065 FOR INSTRUCTIONS IF INTERESTED- 12/28/98 MS13668******************************************** AVAIL. FROM VC536366 VNDR P/N-476-00004 JM/A3040 TRY ADVACOM AS POSSIBLE VENDORLAST PO 1995 OEM JOSLYN CLARK CONTROLS #5M63
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.