GE Healthcare · MRSigna2224666

No signal on Medical Advances Knee coil due to blown fuse in Quick Disconnect box.

Symptom

Auto Prescan Fails With Knee Coil

Error message

2224666 Auto Prescan failed, signal too small. Please try Manual Prescan. Fri Jan 23 14:39:55 2004 2224666 lx-mr psc PscMainControl.cc 1271 Auto Prescan failed, signal too small. Please try Manual Prescan.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2225478-14 Fuse for MedAdv Extremity coil (coil PN is 2225478-8) 2225478-14 Fuse for MedAdv Extremity coil (coil PN is 2225478-8) 2225478-14 Fuse for MedAdv Extremity coil (coil PN is 2225478-8)