GE Healthcare · CTLightSpeed182411
LightSpeed 3.X (Ultra, Plus, QX/i) -Octane
Symptom
Table will not lower below 210 (+-5) mm as shown on Gantry Display. Table will move up and down above 210mm.
Error message
Error: 182411 Motion was prohibited by hardware Interference switches. See long Description for how to DECODE this error Fri Feb 24 21:59:29 2006 Error: 182411 Host: ETC Process: Elevation File: emcControl.c 1.23 11/13/92 Line: 170 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:59:20:810 Log Series: 66 HW/SW error: Motion was prohibited by hardware Interference switches. Hardware Inhibit Switches 0xc0dc
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-136334p23 -Tilt Limit, Cradle Limit, and Tilt Interference Switches. used on HSA, CTi, 1.x LS Gantries and LS 1X - 4X Zeus table See LD for more info 46-136334p23 -Tilt Limit, Cradle Limit, and Tilt Interference Switches. used on HSA, CTi and 1.x LS Part #: 46-136334P23 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: HOME SWITCH-Omron Z series FRU: Yes Status: Orderable Location in Table:
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required