CWA ID85750. Unable to scan with Power Monitor Faults.

Symptom

Unable to scan with Power Monitor Faults

Error message

2254051 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. See Long Description Fri Aug 20 08:33:11 2004 Error: 2254051 Host: lx-mr Process: NSP File: SCP:PmMdsHWO Line: 1019 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 35 pulse width: 33 duty cycle: 09 Fault in head amplitude. This could also occur when power cycling SSM.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required