GE Healthcare · MRExcite2254051
CWA ID85750. Unable to scan with Power Monitor Faults.
Symptom
Unable to scan with Power Monitor Faults
Error message
2254051 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. See Long Description Fri Aug 20 08:33:11 2004 Error: 2254051 Host: lx-mr Process: NSP File: SCP:PmMdsHWO Line: 1019 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 35 pulse width: 33 duty cycle: 09 Fault in head amplitude. This could also occur when power cycling SSM.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.