Pular para o conteúdo principal
GE Healthcare · CTHiSpeed1831101337918329665001

HiSpeed CT/i 4.X with Indigo and HSA Gantry

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183110 KV fault detected: Cathode overcurrent See Long Description for complete error Wed Sep 28 08:31:12 2005 Error: 183110 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 266 Function: XRAY GENERATION : KV Loop Scan: 13379/2/2 Type: Helical Exception Level: Pri/Most Severe Time: 08:31:12:415 Log Series: 332 KV fault detected: Cathode overcurrent. Fault latch address = ffcff3 hex, Bit = D5

Error: 183296 Monitoring: (Measured 245.5 mA vs Ref 240 mA) See Long Description for the full error Error: 183296 Suite: CT01 Host: OBC Process: Generator File: gen_tech.c 1.36 Method: No Method Line: 529 Function: No System Function Reported Scan: 65001/8/38 Type: Axial Exception Level: Sec/Soft Time: 03:11:20:495 Log Series: 49 Selected Technic: 100 kV, 240 mA, 0 % Mod, Backup Timer 4324. VarmA: (Index: Start 0, Skip 0, Last 0) Last Cmd 240 mA. Monitoring: (Measured 245.5 mA vs Ref 240 mA) . 5 Total Samples, Tube Spit Ctr 0. (Rotor 0x0000ae04, Gen 0x9c19b5ff, Mon 0x00020000

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
D3101T / D3102T X-ray Tube Used In CT/i and HiSpeed Advantage Part #: D3102T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TUBE HSA/CTI FRU: Yes Status: Orderable Comments: FLS - NEW 1994 TUBE CAT. FOR RELOAD REPLACES D1199ZR (RELOADED) .9 X .7FS, 1.2-1.2 FS HISPEED ADVANTAGE DIMS 37 X 27 X 28 (LWH) MAXIRAY EQUIVALENT IS D3109T TUBE FAN= 46-219465P1 AM SOURCE = GEMS-MM (MEXICO) *** FLS Tube Names Systems Compatible Heat Capacity Focal Spot Catalog # Model # Insert # Casing # Tube Unit # HSA(Zeus) HiSpeed CT/I, or HiSpeed Advantage 3.5MHu 0.9x0.7/1.2x1.2 D3102T MX165 46-309300G2 46-309500G2 46-309500G801 Tubes started shipping half machined, half cast casings in early 2005. This was stated to be a tempory change in process while a new vendor was being evaluated.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues