GE Healthcare · MRExcite2249150

2249150 T/R driver fault due to bad cable connector attached to head coil.

Symptom

Cannot Scan with Head T/R Driver Error. The system failed to scan either head or body scans. The SSM light indicated T/R Fault.

Error message

2249150 T/R driver fault, head path short. Possibly a shorted diode in the head T/R switch network, or a short circuit load.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2384268 CLASSIC HEAD COIL CBL ASM W/COIL ID See Long Description and Attachment See Attachment for the engineering drawing of this part: 2384268ADW Rev. 2. This is not available online. To confirm this is the most appropriate drawing, go to ePDM.
Area
MR

Full cause & step-by-step resolution