Pular para o conteúdo principal
GE Healthcare · CTHiSpeed18282618287550195

HiSpeed QX/i - Octane

Symptom

Cradle will not drive in and the cradle lock light is on. See the Long Description for more details. Cradle will not drive in and the cradle lock light is on o

Error message

Error: 182826 Communication Error: Communication Packet with the GANTRY DISPLAY. Open Long Description for full error Wed Dec 22 20:49:26 2004 Error: 182826 Host: ETC Process: Gantry Display File: gd_tnc_driver.c Line: 477 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:48:40:485 Log Series: 31 Communication Error: Communication Packet with the GANTRY DISPLAY. Message Code: 2. (Data = 0x20, 0x0, 0x0, 0x0).

Error: 182875 TNC Serial Communications Error During RECEIVING of Data See long description for full error Mon Jul 16 08:35:20 2007 Error: 182875 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 541 Function: No System Function Reported Scan: 50195/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 08:35:20:355 Log Series: 48 TNC Serial Communications Error During RECEIVING of Data. Error Code: 1. Receive Buffer Put Pointer: 17, Receive Buffer Get Pointer: 15. Receive Buffer Count: 2.

Applies to

Models
HiSpeed
Product
HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2372485 ETC IF (Interface or TNC Bd.) Replaces 2229871 See long description for details 2372485 ETC IF (Interface or TNC Bd.) Replaces 2229871 Part #: 2372485 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: ETC INTERFACE W/ AZ60A FRU: Yes Status: Orderable Comments: FLS - ETC INTERFACE W/ AZ60A. ADDED PER PCN 220857. REPLACES P/N 2229871. USE UP EXISTING STK. MS20544 5/1/03. ONE USE LIGHTSPEED PLUS VC:611695-SCI, RAPID CITY, SD CONTROL PCB 46-264368G3
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.