GE Healthcare · CTHiSpeed18880821474157281331218880913592200002358
Scanner won't boot to applications
Symptom
Scanner won't boot to applications. Scanner won't boot to applications
Error message
Error: 188808 Error in RawRead: No space left on device. See LD for Error Details Mon Mar 13 19:15:55 2006 Error: 188808 Host: CT01 Process: RawMgr File: rawDevice.c Line: 259 Error in RawRead: read(10,2147415728,13312) = -1. No space left on device. Error: 188809 Error in _rdh_readRawHeader: RawRead See Long Description for complete error Mon Mar 13 19:15:55 2006 Error: 188809 Host: CT01 Process: RawMgr File: rawfile.c Line: 517 Error in _rdh_readRawHeader: RawRead(header,13592) = 1. Error: 200002358 A HOST_SHUTDOWN was initiatied by the death of RawMgr See LD for details Mon Mar 13 19:15:55 2006 Error: 200002358 Host: CT01 Process: cupMonitor File: ssProcStop.c Line: 998 A HOST_SHUTDOWN was initiatied by the death of RawMgr since other processes depend on it to come up
Applies to
- Models
- HiSpeed
- Product
- HiSpeed YMS Series HiSpeed Series DXi, FXi, LXi, ZXI, and NXI Including Single / Twin detector and Glass / Metal Tubes. See Long Description for Details Subsystem/Components DXI (NP) FXI,LXI (NP+) ZXI (NP++ NXI (NP+ Twin NXI (NP++ Twin OC O2 running Unix or upgraded to Linux PC Gantry Alighnment Lights Halogen Laser Laser Laser Laser Mechanics Common Electronics firmware is different twin Det/Das Twin Det/DAS Table IMS no Yes Yes Yes Yes Generator Jedi Jedi NP++ Jedi Jedi NP++ Jedi Tube D3142T D3142T D3152T D3142T D3152T PDU Common Options List for V/R 5.5, 6, 7, 9, and 10 System 3D Power 440ma Max Image Disk 20000 Smart Prep Fast Scam 0.7 sec Performix 530 cooling Smart Recon Fast Scan 0.8 sec Asymmetric Scan SmartView Fast Scan 1.0 sec Cardiac gating DemtaScan Helical 30 Max Smart Addition Navigator Helical 60 Max Respiratory Gating Helical Plus Helical 120 Max Reformat Cpmmect {rp Raw Data 1000 max VolumeRendering2 ImageDosk Remote Tilt CT Colono Pioneer Mod Enhanced Cooling The following Options name changes Option (V5.5x,V6.0x,V7.x,V9.x,V10x) Option (V8.1x and V11.x) 3D >> Volume Analysis2 Navigator >> Navigator2,VolumeRendering2,VolumeAnalysis2 CtPerfusion >> CTPersusion2 Reformat >> Functool Options V8.1x and V11.x VolumeAnalysis2 Power 440ma ImageDisk 2000 Helical Plus Smart Prep Fast Scan 0.7 sec Performix 530 cooling Connect Pro Smart Recon Fast Scan 0.8 sec Asymmetric Scan ImageDisk SmartView Fast Scan 1.0 sec Cardiac Gating Pioneer MOD* DentaScan Helical 30 Smart Addition Power 200 ma max Navigator2/VolumeRendering2/ VolumeAnalysis Helical 60 Respiratory Gating Power 250 ma Max Power 300 ma Max Helical 120 Thickness 0.5mm CTColon Power 350 ma Max Raw data 1000 CtPerfusion2 AdvVesselAnalysis Remote Tilt Extended FOV Functool Enhance Cooling Performix Tube VolumeRendering2
- Component
- 5166701 (replaces 2376479) SCSI HDD 36GB 3.5IN Recon NP See long description for details Part #: 5166701 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SCSI HDD 36GB 3.5IN Recon NP FRU: Yes Status: Orderable Comments: Part added per ECR2027606 305014118 06/19/2006 Part #: 2376479 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Description: P2011US 36GB 3.5"SCSI HDD FRU: Yes Status: Orderable Comments: revised per to GEHW confirmation ** FE ASSIST ** If no stock of the 2376479 can use 5166701 per ECR 2027606. (212001261 5/9/07)
- Area
- CT
Full cause & step-by-step resolution