Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26013277026010102350444260007402260007453260001512

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260132770 Interface Board Broadcast Communication Error. See long description for a full error example. Fri Jul 13 06:29:48 2007 Error: 260132770 Host: orp Process: orp subsys File: IfbComm.cxx Line: 151 Fri Jul 13 06:29:47 2007 Host : Das Ermes # : 260132770 Exception Class : Abort Severity : Primary File : IfbComm.cxx Line# : 151 Function : No System Function Reported Scan Type : None/Unknown Scan: 873/1/0 Interface Board Broadcast Communication Error.^C See Following Error Message for Digital Interface Board List Command: Acquisition Setup (0x0000135A) Status: Communication timeout - no response (0x00000009).

Do Not Use - use on of the other two Error: 260101023 The (DCB or CCB) has notified the system of an error see long description for examples. Tue Dec 28 09:41:31 2004 Error: 260101023 Host: orp Process: orp_subsys File: rcib_erlog.c Line: 366 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260101023 Exception Class : Abort Severity : Primary File : rcib_erlog.c Line# : 366 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error mess

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5955048 (eqvt part 5151006, 5113710) VCT DIFB Digital Interface Board - Sherlock detector See long description for details. Part #: 5955048 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT DIFB PWA FRU: Yes Status: Orderable Parts Offered as: Current: 5955048 Harvested: Not Available Exchange: Not Available Comments: Per GEC-to-Flextronicw transfer program and ECR 2104697, 212008898, 24 Aug 2010. Old FRU Compatible Yes Comments Per GEC to Flextronics Transfer Program #020000, 212008898, 8 Sept 2010. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5955048 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number BOK39368 Old FRU Compatible Yes Comments Per GEC to Flextronics Transfer Program #020000, 212008898, 8 Sept 2010. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5151006 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number BOK39368 5151006 (eqvt part 5113710) VCT DIFB Digital Interface Board - Sherlock detector Part #: 5151006 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT DIFB PWA FRU: Yes Status: Orderable ** FE ASSIST ** If no stock can use P/N 5113710 (212001261 10/12/07) Part #: 5113710 Part Type: GP CONSUMABLES Part Weight: -90719.09 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT DIFB FRU: Yes Status: Orderable ** FE ASSIST ** If no stock can use P/N 5151006 (212001261 10/12/07)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.